- Waters ACQUITY UPLC I-Class PLUS System
- Q-Orbitrap Exploris 240
- Metabolomics Services
- High-Performance Liquid Chromatography (HPLC)
- Autoflex speed™ MALDI TOF/TOF
- Enhanced Speed with a 2 kHz (kiloHertz) proprietary smartbeam-II solid-state laser
(modified Nd:YAG laser) 355 nm and electronics for high throughput applications
including MALDI Imaging, and high-throughput microbiology ID
• Laser focus diameters down to 10 µm for high spatial resolution imaging without
pixel overlap.
• TOF-analyzer for linear and reflectron mode, MS/MS capability TOF-analyzer for
both positive and negative ion mode
• Reflectron resolution ≥ 26000 FWHM ( Somatostatin 28; m/z 3147.47 Da), mass
accuracy < 10ppm (external calibration)
• Linear resolution ≥1,100 for Cytochrome C (m/z 12,361) and ≥ 600 for Protein
A (m/z 44,613) . linear mass accuracy for Cytochrome C <100 ppm (external
calibration)
• Mass accuracy (protein mixture): with external calibration: better than 100 ppm
with internal calibration: better than 90 ppm
• MS sensitivity: 500 fmol BSA (m/z 66,000). S/N ≥ 100
• Expanded mass range capabilities to handle intact proteins and molecules such
as large protein and polymers.
LIFT™ – the latest TOF/TOF technology to enable the use of various MS/MS
techniques, including LID and high-energy CID
• Minimized maintenance with IR laser-based self-cleaning MALDI Perpetual™ Ion
Source, which improves system up-time and utility.
• Optimized Software packages for numerous applications e.g. Protein-, Peptide-,
Polymer-Analyses, and MALDI Imaging, clinical research methods and the
MALDI Biotyper workflow.Location: Building 51, Room 232
- Q-Exactive/Q-Exactive Plus
- Discovery Proteomics: Handles foundational screening by consistently identifying thousands of proteins from complex biological samples.
- Quantitative Analysis: Processes label-free and TMT-labeled samples for precise protein quantification.
- PTM Analysis: Detects and localizes complex post-translational modifications.
- Resolving power: up to 140,000 Full Width at Half Maximum (FWHM).
- Mass Range for Full Scans: m/z 50–6000 (Scan range: last mass ≤ 15 × first mass), and mass accuracy is 3ppm.
Q-Exactive Location: Building 51, Room 230
Q-Exactive Plus Location: Building 51, Room 232
- Gas Chromatography/Mass Spectrometry (GC/MS)
The Proteomics Unit services include:
• Consultation in experiment design
• Sample preparation and method development
• LCMS method development and optimization
• Data acquisition (DDA, DIA, PRM)
• Data analysis
• Down-stream support in preparation to publications
• Training
High-precision binary pump with an adaptable sample manager for vials and plates and an extensive injection volume range. The autosampler has a refrigerated chamber and high-precision needle for stable, high-throughput workflows.
The next generation of high-resolution mass-spectrometers that can be used for Data Independent Analysis (DIA) and Parallel Reaction Monitoring (PRM) with a higher scan speed and faster polarity switch. The MS is equipped with an ESI source that provides soft ionization of the molecules. Resolving power is up to 240,000 Full Width at Half Maximum (FWHM). Mass Range for Full Scans is m/z 40–3000 (Scan range: last mass ≤ 15 × first mass), and mass accuracy is 1ppm.
The Unit services include:
• Consultation in experiment design
• Assistance with sample preparation method development
• UPLC-MS method development and optimization
• Data acquisition
• Basic and advanced data analysis
• Down-stream support in preparation to publications
• Training
High-Performance Liquid Chromatography (HPLC) is used for the qualitative and quantitative analysis of organic molecules and organic materials that are not sufficiently volatile for GC analysis (including high–molecular-weight molecules). Laboratory`s HPLC Agilent 1100 Series is equipped with photodiode array and fluorescence detectors.
Wavelength range: 190-950 nm
Matrix-Assisted Laser Desorption/Ionization(MALDI) stands as one of the most powerful
techniques invented for investigating:
1. Biopolymers (such as peptides, proteins, glycansand oligonucleotides)
2. Synthetic polymers and organic molecules (>500 Da), especially in mixtures and crude samples
3. MALDI Imaging. Distribution of proteins,lipid, smallmolecules and etc. in tissue sections.
Results are shown as color coded intensity plots and can be overlaid to an image of the
investigated sample
5. TLC MALDI. Measurements directly from a thin layerchromatography (TLC)
6. Microorganism and Fungi identification and classification
7. Detailed sample characterization by MS/MS.
8.Top down proteomics.
The Thermo Scientific Q Exactive mass spectrometer serves the proteomics unit, delivering the steady performance required for routine operations. Its combination of a quadrupole and an Orbitrap mass analyzer continue to provide the high resolution and mass accuracy necessary for everyday workflows. The instrument primarily handles:
High-resolution mass-spectrometers that can be used for both Data Independent Analysis (DIA) and Parallel Reaction Monitoring (PRM). The MS is equipped with an ESI source that provides soft ionization of the molecules.
Gas Chromatography/Mass Spectrometry (GC/MS) is applied for the separation and identification of a variety of organic compounds. The Unit’s Agilent 7890N GC is connected to an Agilent 5977A single-quadrupole mass-selective detector. The instrument is equipped with a 100-vial autosampler, an NIST02 MS Library (includes 175,000 spectra for 147,000 compounds) and an ACD Labs MS Manager (software package for mass-spectra interpretation and structure elucidation).
Performance electronics for 12,500 u/s scan speed (8,000 u/s write-to-disk)