Autoflex speed™ MALDI TOF/TOF

Matrix-Assisted Laser Desorption/Ionization(MALDI) stands as one of the most powerful
techniques invented for investigating:

1. Biopolymers (such as peptides, proteins, glycansand oligonucleotides)

2. Synthetic polymers and organic molecules (>500 Da), especially in mixtures and crude samples

3. MALDI Imaging. Distribution of proteins,lipid, smallmolecules and etc. in tissue sections.
Results are shown as color coded intensity plots and can be overlaid to an image of the
investigated sample

5. TLC MALDI. Measurements directly from a thin layerchromatography (TLC)

6. Microorganism and Fungi identification and classification

7. Detailed sample characterization by MS/MS.

8.Top down proteomics.

Specifications:
  • Enhanced Speed with a 2 kHz (kiloHertz) proprietary smartbeam-II solid-state laser
    (modified Nd:YAG laser) 355 nm and electronics for high throughput applications
    including MALDI Imaging, and high-throughput microbiology ID
    • Laser focus diameters down to 10 µm for high spatial resolution imaging without
    pixel overlap.
    TOF-analyzer for linear and reflectron mode, MS/MS capability TOF-analyzer for
    both positive and negative ion mode
    Reflectron resolution ≥ 26000 FWHM ( Somatostatin 28; m/z 3147.47 Da), mass
    accuracy < 10ppm (external calibration)
    Linear resolution ≥1,100 for Cytochrome C (m/z 12,361) and ≥ 600 for Protein
    A (m/z 44,613) . linear mass accuracy for Cytochrome C <100 ppm (external
    calibration)
    Mass accuracy (protein mixture): with external calibration: better than 100 ppm
    with internal calibration: better than 90 ppm
    • MS sensitivity: 500 fmol BSA (m/z 66,000). S/N ≥ 100
    Expanded mass range capabilities to handle intact proteins and molecules such
    as large protein and polymers.
    LIFT™ – the latest TOF/TOF technology to enable the use of various MS/MS
    techniques, including LID and high-energy CID
    Minimized maintenance with IR laser-based self-cleaning MALDI Perpetual™ Ion
    Source, which improves system up-time and utility.
    Optimized Software packages for numerous applications e.g. Protein-, Peptide-,
    Polymer-Analyses, and MALDI Imaging, clinical research methods and the
    MALDI Biotyper workflow.

    Location: Building 51, Room 232